Cause and Treatment of Spray Failure in Salt Spray Test Chamber

July 13, 2020

Latest company news about Cause and Treatment of Spray Failure in Salt Spray Test Chamber

Insufficient Spray Volume:
1. The spray regulator is placed too low, adjust the spray regulator to high
2. The glass filter in the preheating tank is blocked, and the glass filter is cleaned
3. Check whether the nozzle is partially blocked, and if it is blocked, please discharge the debris
4. The pressure setting is too low, adjust the pressure regulating valve


Unable to Spray:
1. The air compressor is not running, turn on the air compressor button, check whether the air pipe link of the air compressor is correct
2. Check whether the pressure of the pressure regulating valve is turned on, please adjust to the normal pressure value
3.Check whether the nozzle is blocked, the nozzle is a consumable item, generally more than 1500 hours of use is normal damage, the nozzle should be replaced
4. Check whether the spray switch is damaged, if it is damaged, please replace the switch
5. Check whether the spray solenoid valve is damaged. If there is intake air but no air outlet, remove the solenoid valve and clean the solenoid valve. If the coil is burned, it is necessary to replace the matching solenoid valve coil, otherwise the entire solenoid valve is replaced
6. Check whether there is salt water, no salt water, add appropriate amount of salt water